TY - JOUR
T1 - Five-Year Outcomes in Patients with Fully Magnetically Levitated vs Axial-Flow Left Ventricular Assist Devices in the MOMENTUM 3 Randomized Trial
AU - Mehra, Mandeep R.
AU - Goldstein, Daniel J.
AU - Cleveland, Joseph C.
AU - Cowger, Jennifer A.
AU - Hall, Shelley
AU - Salerno, Christopher T.
AU - Naka, Yoshifumi
AU - Horstmanshof, Douglas
AU - Chuang, Joyce
AU - Wang, Aijia
AU - Uriel, Nir
N1 - Funding Information:
Funding/Support: This study was sponsored by
Publisher Copyright:
© 2022 American Medical Association. All rights reserved.
PY - 2022/9/27
Y1 - 2022/9/27
N2 - Importance: Although durable left ventricular assist device (LVAD) therapy has emerged as an important treatment option for patients with advanced heart failure refractory to pharmacological support, outcomes, including survival, beyond 2 years remain poorly characterized. Objective: To report the composite end point of survival to transplant, recovery, or LVAD support free of debilitating stroke (Modified Rankin Scale score >3) or reoperation to replace the pump 5 years after the implant in participants who received the fully magnetically levitated centrifugal-flow HeartMate 3 or axial-flow HeartMate II LVAD in the MOMENTUM 3 randomized trial and were still receiving LVAD therapy at the 2-year follow-up. Design, Setting, and Participants: This observational study was a 5-year follow-up of the MOMENTUM 3 trial, conducted in 69 US centers, that demonstrated superiority of the centrifugal-flow LVAD to the axial-flow pump with respect to survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump at 2 years. A total of 295 patients were enrolled between June 2019 to April 2021 in the extended-phase study, with 5-year follow-up completed in September 2021. Exposures: Of 1020 patients in the investigational device exemption per-protocol population, 536 were still receiving LVAD support at 2 years, of whom 289 received the centrifugal-flow pump and 247 received the axial-flow pump. Main Outcomes and Measures: There were 10 end points evaluated at 5 years in the per-protocol population, including a composite of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump between the centrifugal-flow and axial-flow pump groups and overall survival between the 2 groups. Results: A total of 477 patients (295 enrolled and 182 provided limited data) of 536 patients still receiving LVAD support at 2 years contributed to the extended-phase analysis (median age, 62 y; 86 [18%] women). The 5-year Kaplan-Meier estimate of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump in the centrifugal-flow vs axial-flow group was 54.0% vs 29.7% (hazard ratio, 0.55 [95% CI, 0.45-0.67]; P <.001). Overall Kaplan-Meier survival was 58.4% in the centrifugal-flow group vs 43.7% in the axial-flow group (hazard ratio, 0.72 [95% CI, 0.58-0.89]; P =.003). Serious adverse events of stroke, bleeding, and pump thrombosis were less frequent in the centrifugal-flow pump group. Conclusions and Relevance: In this observational follow-up study of patients from the MOMENTUM 3 randomized trial, per-protocol analyses found that receipt of a fully magnetically levitated centrifugal-flow LVAD vs axial-flow LVAD was associated with a better composite outcome and higher likelihood of overall survival at 5 years. These findings support the use of the fully magnetically levitated LVAD. Trial Registration: ClinicalTrials.gov Identifier: NCT02224755 and NCT03982979.
AB - Importance: Although durable left ventricular assist device (LVAD) therapy has emerged as an important treatment option for patients with advanced heart failure refractory to pharmacological support, outcomes, including survival, beyond 2 years remain poorly characterized. Objective: To report the composite end point of survival to transplant, recovery, or LVAD support free of debilitating stroke (Modified Rankin Scale score >3) or reoperation to replace the pump 5 years after the implant in participants who received the fully magnetically levitated centrifugal-flow HeartMate 3 or axial-flow HeartMate II LVAD in the MOMENTUM 3 randomized trial and were still receiving LVAD therapy at the 2-year follow-up. Design, Setting, and Participants: This observational study was a 5-year follow-up of the MOMENTUM 3 trial, conducted in 69 US centers, that demonstrated superiority of the centrifugal-flow LVAD to the axial-flow pump with respect to survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump at 2 years. A total of 295 patients were enrolled between June 2019 to April 2021 in the extended-phase study, with 5-year follow-up completed in September 2021. Exposures: Of 1020 patients in the investigational device exemption per-protocol population, 536 were still receiving LVAD support at 2 years, of whom 289 received the centrifugal-flow pump and 247 received the axial-flow pump. Main Outcomes and Measures: There were 10 end points evaluated at 5 years in the per-protocol population, including a composite of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump between the centrifugal-flow and axial-flow pump groups and overall survival between the 2 groups. Results: A total of 477 patients (295 enrolled and 182 provided limited data) of 536 patients still receiving LVAD support at 2 years contributed to the extended-phase analysis (median age, 62 y; 86 [18%] women). The 5-year Kaplan-Meier estimate of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation to replace the pump in the centrifugal-flow vs axial-flow group was 54.0% vs 29.7% (hazard ratio, 0.55 [95% CI, 0.45-0.67]; P <.001). Overall Kaplan-Meier survival was 58.4% in the centrifugal-flow group vs 43.7% in the axial-flow group (hazard ratio, 0.72 [95% CI, 0.58-0.89]; P =.003). Serious adverse events of stroke, bleeding, and pump thrombosis were less frequent in the centrifugal-flow pump group. Conclusions and Relevance: In this observational follow-up study of patients from the MOMENTUM 3 randomized trial, per-protocol analyses found that receipt of a fully magnetically levitated centrifugal-flow LVAD vs axial-flow LVAD was associated with a better composite outcome and higher likelihood of overall survival at 5 years. These findings support the use of the fully magnetically levitated LVAD. Trial Registration: ClinicalTrials.gov Identifier: NCT02224755 and NCT03982979.
UR - http://www.scopus.com/inward/record.url?scp=85138150184&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85138150184&partnerID=8YFLogxK
U2 - 10.1001/jama.2022.16197
DO - 10.1001/jama.2022.16197
M3 - Article
C2 - 36074476
AN - SCOPUS:85138150184
SN - 0098-7484
VL - 328
SP - 1233
EP - 1242
JO - JAMA - Journal of the American Medical Association
JF - JAMA - Journal of the American Medical Association
IS - 12
ER -