Abstract
In monitoring a patient with chronic pain who was taking high-dose morphine and oxycodone with weekly urine enzymatic immunoassay (EIA) toxicology testing, the authors noted consistent positives for buprenorphine. The patient was not taking buprenorphine, and gas chromatography/mass spectroscopy (GCMS) testing on multiple samples revealed no buprenorphine, indicating a case of false-positive buprenorphine EIAs in a high-dose opiate case. The authors discontinued oxycodone for a period of time and then discontinued morphine. Urine monitoring with EIAs and GCMS revealed false-positive buprenorphine EIAs, which remained only when the patient was taking morphine. When taking only oxycodone and no morphine, urine samples became buprenorphine negative. When morphine was reintroduced, false-positive buprenorphine results resumed. Medical practitioners should be aware that high-dose morphine (with morphine urine levels turning positive within the15,000 to 28,000 mg/mL range) may produce false-positive buprenorphine EIAs with standard urine EIA toxicology testing.