Abstract
Case: A 52-year-old man who previously had undergone total hip arthroplasty with use of an Epoch (Zimmer) extensively porous-coated femoral stem presented with hip pain after a high-energy motor-vehicle accident. Radiographs demonstrated delamination at the polyaryletheretherketone-cobalt-chromium (PEEK-CoCr) core interface, which led to loosening. At revision, the implant demonstrated gross evidence of debonding at the PEEK-CoCr interface. Conclusion: We demonstrate clinical and tribologic evidence of a rare failure mode involving the PEEK-metal interface in a composite femoral stem, which surgeons should be aware of when evaluating patients with this type of implant.
Original language | English (US) |
---|---|
Article number | e83 |
Journal | JBJS Case Connector |
Volume | 7 |
Issue number | 4 |
DOIs | |
State | Published - Jan 1 2017 |
Externally published | Yes |
ASJC Scopus subject areas
- Surgery
- Orthopedics and Sports Medicine