TY - JOUR
T1 - Complications of Spinal Cord Stimulators—A Comprehensive Review Article
AU - Koushik, Sarang S.
AU - Raghavan, Jagun
AU - Saranathan, Shreya
AU - Slinchenkova, Kateryna
AU - Viswanath, Omar
AU - Shaparin, Naum
N1 - Publisher Copyright:
© The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature 2023.
PY - 2024/1
Y1 - 2024/1
N2 - Purpose of Review: Spinal cord stimulation has been increasing in influence as an option to regulate pain, especially in the chronic pain patient population. However, even with the numerous changes made to this technology since its inception, it is still prone to various complications such as hardware issues, neurological injury/epidural hematoma, infections, and other biological concerns. The purpose of this article is to thoroughly review and evaluate literature pertaining to the complications associated with percutaneous spinal cord stimulation. Recent Findings: Lead migration is generally the most common complication of percutaneous spinal cord stimulation; however, recent utilization of various anchoring techniques has been discussed and experienced clinical success in decreasing the prevalence of lead migration and lead fractures. With newer high-frequency systems gaining traction to improve pain management and decrease complications as compared to traditional systems, rechargeable implantable pulse generators have been the preferred power source. However, recent findings may suggest that these rechargeable implantable pulse generators do not significantly increase battery life as much as was proposed. Intraoperative neuromonitoring has seen success in mitigating neurological injury postoperatively and may see more usage in the future through more testing. Though the occurrence of infection and biological complications, including dural puncture and skin erosion, has been less frequent over time, they should still be treated in accordance with established protocols. Summary: While many complications can arise following percutaneous spinal cord stimulator implantation, the procedure is less invasive than open implantation and has seen largely positive patient feedback. Hardware complications, the more common issues that can occur, rarely indicate a serious risk and can generally be remedied through reoperation. However, less common cases such as neurological injury, infections, and biological complications require prompt diagnosis to improve the condition of the patient and prevent significant damage.
AB - Purpose of Review: Spinal cord stimulation has been increasing in influence as an option to regulate pain, especially in the chronic pain patient population. However, even with the numerous changes made to this technology since its inception, it is still prone to various complications such as hardware issues, neurological injury/epidural hematoma, infections, and other biological concerns. The purpose of this article is to thoroughly review and evaluate literature pertaining to the complications associated with percutaneous spinal cord stimulation. Recent Findings: Lead migration is generally the most common complication of percutaneous spinal cord stimulation; however, recent utilization of various anchoring techniques has been discussed and experienced clinical success in decreasing the prevalence of lead migration and lead fractures. With newer high-frequency systems gaining traction to improve pain management and decrease complications as compared to traditional systems, rechargeable implantable pulse generators have been the preferred power source. However, recent findings may suggest that these rechargeable implantable pulse generators do not significantly increase battery life as much as was proposed. Intraoperative neuromonitoring has seen success in mitigating neurological injury postoperatively and may see more usage in the future through more testing. Though the occurrence of infection and biological complications, including dural puncture and skin erosion, has been less frequent over time, they should still be treated in accordance with established protocols. Summary: While many complications can arise following percutaneous spinal cord stimulator implantation, the procedure is less invasive than open implantation and has seen largely positive patient feedback. Hardware complications, the more common issues that can occur, rarely indicate a serious risk and can generally be remedied through reoperation. However, less common cases such as neurological injury, infections, and biological complications require prompt diagnosis to improve the condition of the patient and prevent significant damage.
KW - Dural puncture
KW - Epidural hematoma
KW - Hardware complications
KW - Infection
KW - Lead migration
KW - Percutaneous spinal cord stimulation
UR - http://www.scopus.com/inward/record.url?scp=85174486470&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85174486470&partnerID=8YFLogxK
U2 - 10.1007/s11916-023-01178-3
DO - 10.1007/s11916-023-01178-3
M3 - Review article
C2 - 37855944
AN - SCOPUS:85174486470
SN - 1531-3433
VL - 28
SP - 1
EP - 9
JO - Current pain and headache reports
JF - Current pain and headache reports
IS - 1