Abstract
To determine the accuracy of Fourier filtration in removing the high-frequency component of noise from acoustic quantification (AQ) echocardiographic images, we processed 800 parasternal short-axis images obtained from 10 study subjects. M-mode tracings were also obtained and used as gold standard for correlating the results from raw AQ and Fourier-filtered images. Left ventricular short-axis diameters measured from the raw AQ and Fourier-filtered data were compared with the M-mode diameters (r = 0.91, p < 0.001 for raw AQ; and r = 0.96, p < 0.001, for Fourier filtered images). Fractional shortening showed better correlation between Fourier-filtered images and M-mode (r -0.79, p < 0.03) versus raw AQ and M-mode (r = 0.33, p = 0.46). Beat-to-beat reproducibility was also found to be better for fractional area change (r = 0.82, p = 0.01 versus r = 0.66, p = 0.07), peak area filling rate (r = 0.87, p = 0.004 versus r = 0.62, p = 0.1), and peak area emptying rate (r = 0.99, p < 0.0001 versus r = 0.19, p < 0.7) for Fourier filtered versus raw AQ. Our results indicate that Fourier filtration of AQ data results in more accurate renresenfation of the true endocardial borders.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 310-319 |
| Number of pages | 10 |
| Journal | Journal of the American Society of Echocardiography |
| Volume | 10 |
| Issue number | 4 |
| DOIs | |
| State | Published - 1997 |
| Externally published | Yes |
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging
- Cardiology and Cardiovascular Medicine
Fingerprint
Dive into the research topics of 'Two-dimensional fourier filtration of acoustic quantification echocardiographic images: Improved reproducibility and accuracy of automated measurements of left ventricular performance'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS