Abstract
Here, we retrospectively analyzed the comparative results of 182 paired dry nasopharyngeal swabs tested by Abbott ID NOW and nasopharyngeal swabs in viral transport medium by real-time RT-PCR methods. While the overall agreement was 96.2%, we found that of 15 samples that were tested positive with RT-PCR methods, 7 were missed by ID NOW, resulting in a false-negative rate of 47%.
| Original language | English (US) |
|---|---|
| Article number | 115123 |
| Journal | Diagnostic Microbiology and Infectious Disease |
| Volume | 98 |
| Issue number | 2 |
| DOIs | |
| State | Published - Oct 2020 |
| Externally published | Yes |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- COVID-19, Abbott ID NOW
- Dry nasopharyngeal swab
- SARS-CoV-2
ASJC Scopus subject areas
- Microbiology (medical)
- Infectious Diseases
Fingerprint
Dive into the research topics of 'How many are we missing with ID NOW COVID-19 assay using direct nasopharyngeal swabs? Findings from a mid-sized academic hospital clinical microbiology laboratory'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS