Abstract
Implantation of an A V universal (DDD) pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS. (PACE, Vol. 5. July‐August, 1982)
| Original language | English (US) |
|---|---|
| Pages (from-to) | 486-489 |
| Number of pages | 4 |
| Journal | Pacing and Clinical Electrophysiology |
| Volume | 5 |
| Issue number | 4 |
| DOIs | |
| State | Published - Jul 1982 |