Abstract

Background: With transcatheter aortic valve implantation (TAVI) now extending to lower-risk and younger patients, optimizing procedural and hemodynamic outcomes is critical. The Myval Octacor, a new balloon-expandable valve (BEV), was developed to improve outcomes by reducing paravalvular regurgitation (PVL), minimizing pacemaker implantation (PPI) rates, and enhancing hemodynamic performance. However, limited data are available comparing Myval Octacor to contemporary self-expanding supra-annular valves (SEVs) Evolut PRO/PRO+ and Acurate Neo2. Aims: This study aimed to compare the safety, efficacy, and short-term clinical outcomes of Myval Octacor with SEVs in transfemoral TAVI patients. Methods: We conducted a prospective, multicenter registry including patients treated with Myval Octacor, then compared them to SEV recipients from the NEOPRO2 registry. Propensity score matching adjusted for baseline differences between groups. The primary endpoint was 30-day Valve Academic Research Consortium-3 (VARC-3) device success. Secondary endpoints included technical success, valve performance, and early safety outcomes. Results: Among 252 Myval Octacor patients and 2175 SEV patients, 90 matched pairs were compared. Myval Octacor patients showed higher 30-day VARC-3 device success than SEVs patients (97% vs. 88%, p = 0.024), primarily due to a numerically lower rate of moderate-to-severe PVL (1% vs. 7%, p = 0.06). The Octacor group also exhibited larger indexed effective orifice areas. Mortality, stroke, PPI, and myocardial infarction rates were similar between groups at 30 days. Conclusions: The Myval Octacor demonstrated comparable early safety and efficacy to supra-annular SEVs, with advantages in device success rate and hemodynamic performance. Adequately sized randomized study is required to confirm these findings.

Original languageEnglish (US)
Pages (from-to)2949-2957
Number of pages9
JournalCatheterization and Cardiovascular Interventions
Volume106
Issue number5
DOIs
StatePublished - Nov 1 2025

Keywords

  • aortic stenosis
  • balloon-expandable Myval
  • self-expanding Acurate Neo2
  • self-expanding Evolut PRO/PRO+
  • transcatheter aortic valve replacement
  • transcatheter heart valves

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Balloon-Expandable Myval Octacor Versus Self-Expanding Evolut PRO/PRO+ and Acurate Neo2: Short-Term Outcomes From Propensity-Matched Analysis'. Together they form a unique fingerprint.

Cite this